Repair the DATABASE
For a repairing of a database you can doe the following steps:
All in the same system:
1.Your system have to stay in the Main Menu, not in any programs
2.Create a new Folder C:Datayymmdd (for Example C:Data020625)
Make a copy from C: RETI RETIBASE RETIBASE to the new folder C: Datayymmdd
Example C:Data020625 ( here are all files )
Now you have got a copy of all data on C: in the new folder. This is better; if you have any problems, you have all data on this place.
3.Delete all files in the Folder C: RETI RETIBASE RETIBASE.CPY
4.Copy all Files from C: RETI RETIBASE RETIBASE to C: RETI RETIBASE RETIBASE.CPY
5.Now you can delete the Folder C: RETI RETIBASE RETIBASE
Important: Before delete, you have to have the copy in C: Datayymmdd and in C: RETI RETIBASE RETIBASE.CPY
6.Now start the RETI-scan program - The system gives the message: database are damage, create a new database - please answer yes. Now you have a new database without data. In next steps you have to import all old DATA.
7.Select in the Program RETIscan: DATABASE Maintance and select in this the option: Import DATABASE from Drive C: ( The system imports the data from RETIBASE.CPY )
Now all Patient Data and all (only ) RETISCAN Tests are imported in the new DATABASE.
To get the RETIport DATA in the new DATABASE, please start the RETI-port program and do the same procedure. Than all Patient Data and all VEP+ERG ( only ) Tests are imported in the new DATABASE.
For new EOG and Nystagmography Versions, please do the same procedure in each program and you will get all data in the new DATABASE step by step.
Repair the DATABASEFor a repairing of a database you can doe the following steps:All in the same system:1.Your system have to stay in the Main Menu, not in any programs2.Create a new Folder C:Datayymmdd (for Example C:Data020625)Make a copy from C: RETI RETIBASE RETIBASE to the new folder C: DatayymmddExample C:Data020625 ( here are all files ) Now you have got a copy of all data on C: in the new folder. This is better; if you have any problems, you have all data on this place.3.Delete all files in the Folder C: RETI RETIBASE RETIBASE.CPY4.Copy all Files from C: RETI RETIBASE RETIBASE to C: RETI RETIBASE RETIBASE.CPY 5.Now you can delete the Folder C: RETI RETIBASE RETIBASEImportant: Before delete, you have to have the copy in C: Datayymmdd and in C: RETI RETIBASE RETIBASE.CPY6.Now start the RETI-scan program - The system gives the message: database are damage, create a new database - please answer yes. Now you have a new database without data. In next steps you have to import all old DATA.7.Select in the Program RETIscan: DATABASE Maintance and select in this the option: Import DATABASE from Drive C: ( The system imports the data from RETIBASE.CPY )Now all Patient Data and all (only ) RETISCAN Tests are imported in the new DATABASE.To get the RETIport DATA in the new DATABASE, please start the RETI-port program and do the same procedure. Than all Patient Data and all VEP+ERG ( only ) Tests are imported in the new DATABASE.For new EOG and Nystagmography Versions, please do the same procedure in each program and you will get all data in the new DATABASE step by step.