Reflectance Calibration Procedure
1. Touch and release Cell 5 on the Main Menu Display until the desired specular component
status (SCE/SCI) is displayed.
2. Select the desired sampling aperture. Refer to “Changing the Sampling Aperture” later in this
section.
3. Use the software program to select reflectance mode.
4. Place the Zero Calibration Standard facing the unit so that it completely covers the viewport
opening.
5. Use the software program to initiate the zero (black) calibration.
6. Place the White Calibration Tile facing the unit so that it completely covers the viewport
opening.
7. Use the software program to initiate the white tile calibration.
8. After calibration return the white calibration tile to it storage area.