For resuscitators designated for use with a body mass less than 10 kg, a pressure-limiting system shall be provided so that the airway pressure does not exceed 4,5 kPa (-45 cmH20) under the test conditions described in A.4.12. However, it shall be possible to generate an airway pressure of at least 3 kPa (30 cm H20).