Note:
roubleshooting of Assay Performance Check
We highly recommend that the high voltage is not changed in the first step if the signal level
of the BCR2 mean is outside the range of ± 7% of the target value.
First of all, the measuring cell preparation should be extended by another 30 or 50
measuring cell preparation cycles (refer to Chapter 2.3.9). In the second step make
sure that the measuring cell is mounted properly and that the BlankCell reagent is
O.K. Refer also to Chapters T