The 5.3 intermediate precision test: according to "5.2" under the method for the preparation of test sample solution, by different analysts using different instruments were measured, respectively, into the sample for 6 times, the content of RSD value calculation.
6 accuracy (recovery test): the sample of half amount, at the same time, according to the determination results of 80%, 100%, 120% add the reference solution (3 copies of each concentration), for recovery of the added sample test, recovery rate should be 95.0%~105.0%, 9 copies of the RSD should be no more than 2%.
7 tests of solution stability were investigated: a reference solution and the stability of the test solution at room temperature stability and a refrigerator,As the reference solution to save time and sample provides the basis for the determination of time. The durability test of
8 chromatographic conditions: the effects of different column temperature (- 5 DEG C), different flow rates (+ 10%), different chromatographic column, different instruments, different detection wavelengths on determination results and application effects of the system of the situation (tailing factor, the degree of separation, the number of theoretical plates) to determine the effects of chromatographic conditions, the range of acceptance.
9 the limit: the determination results of samples of different production date (the 10 batch) and the stability of the test results of 3 batches of sample to determine the content limit.