6.2 抗弯折性能
测试项目及弯折次数按表4规定。
7.耐清洗消毒性能
正常使用时与患者接触的部件,应能承受在正常使用时可能遇到的或由制造厂在使用说明书中规定的清洗,消毒和灭菌,而又不损坏或影响其安全防护性能和使用性能。

8.生物学评价
探头与人体皮肤接触的部件,应符合以下要求:
a) 细胞毒性试验,应不大于1级;
b) 迟发型超敏反应试验,应无致敏反应;
c) 皮肤刺激试验,应不大于极轻微刺激反应。

9.耐灭菌性(可重复使用产品适用)
经环氧乙烷反复灭菌10次后,应符合4.2、4.3的要求。
0/5000
原始語言: -
目標語言: -
結果 (英文) 1: [復制]
復制成功!
6.2 bending resistance performanceTable 4 requirements the number of test items and bending.7. resistance to cleaning and disinfection performanceParts in contact with the patient when used properly, you may encounter should be able to withstand normal use or by the manufacturer in the instruction manual provided for the cleaning, disinfection and sterilization, without damage or affect its performance and safety performance.8. biological evaluationSensor parts in contact with human skin, shall conform to the following requirements: A) cytotoxicity test should not be greater than 1;B) delayed-type hypersensitivity test should be no allergic reaction;C) skin irritation test shall not be more than minor irritation.9. sterilization (reusable products)After repeated sterilization by ethylene oxide 10, should meet the requirements of 4.2 and 4.3.
正在翻譯中..
結果 (英文) 2:[復制]
復制成功!
6.2 kink performance
test items and number of bending according to Table 4.
7. resistance to cleaning and disinfection performance
of normal use in contact with patients member should be able to withstand in normal use may encounter or cleaning, disinfection and sterilization by the manufacturer in the instructions for use specified, without damaging or affecting its safety properties and performance.

8. Biological Evaluation
probe member in contact with human skin, shall meet the following requirements:
A) cytotoxicity test, should be no greater than grade 1;
B) delayed-type hypersensitivity test, should be no sensitization;
C) skin irritation test It should be no more than very minor irritation.

9. sterilization resistance (reusable products applicable)
by ethylene oxide sterilization repeatedly 10 times, 4.2, 4.3 should meet the requirements of.
正在翻譯中..
結果 (英文) 3:[復制]
復制成功!
6.2 bending resistanceTest items and the number of times of bending according to table 4.7 cleaning and disinfection performanceNormal use and patient contact parts, should be able to withstand that may be encountered during normal use or prescribed by the manufacturer in the instructions for use in cleaning, disinfection and sterilization, and do not damage or affect its safety performance and service performance.8 biological evaluationProbes and parts of human skin contact, should meet the following requirements:A) cell toxicity test, should not be more than 1;B) delayed type hypersensitivity test, there should be no allergic reaction;C) skin irritation test, should not be greater than the very slight stimulation reaction.9 sterilization resistance (reusable product application)After repeated sterilization by ethylene oxide 10 times, should meet the requirements of 4.3, 4.2.
正在翻譯中..
 
其它語言
本翻譯工具支援: 世界語, 中文, 丹麥文, 亞塞拜然文, 亞美尼亞文, 伊博文, 俄文, 保加利亞文, 信德文, 偵測語言, 優魯巴文, 克林貢語, 克羅埃西亞文, 冰島文, 加泰羅尼亞文, 加里西亞文, 匈牙利文, 南非柯薩文, 南非祖魯文, 卡納達文, 印尼巽他文, 印尼文, 印度古哈拉地文, 印度文, 吉爾吉斯文, 哈薩克文, 喬治亞文, 土庫曼文, 土耳其文, 塔吉克文, 塞爾維亞文, 夏威夷文, 奇切瓦文, 威爾斯文, 孟加拉文, 宿霧文, 寮文, 尼泊爾文, 巴斯克文, 布爾文, 希伯來文, 希臘文, 帕施圖文, 庫德文, 弗利然文, 德文, 意第緒文, 愛沙尼亞文, 愛爾蘭文, 拉丁文, 拉脫維亞文, 挪威文, 捷克文, 斯洛伐克文, 斯洛維尼亞文, 斯瓦希里文, 旁遮普文, 日文, 歐利亞文 (奧里雅文), 毛利文, 法文, 波士尼亞文, 波斯文, 波蘭文, 泰文, 泰盧固文, 泰米爾文, 海地克里奧文, 烏克蘭文, 烏爾都文, 烏茲別克文, 爪哇文, 瑞典文, 瑟索托文, 白俄羅斯文, 盧安達文, 盧森堡文, 科西嘉文, 立陶宛文, 索馬里文, 紹納文, 維吾爾文, 緬甸文, 繁體中文, 羅馬尼亞文, 義大利文, 芬蘭文, 苗文, 英文, 荷蘭文, 菲律賓文, 葡萄牙文, 蒙古文, 薩摩亞文, 蘇格蘭的蓋爾文, 西班牙文, 豪沙文, 越南文, 錫蘭文, 阿姆哈拉文, 阿拉伯文, 阿爾巴尼亞文, 韃靼文, 韓文, 馬來文, 馬其頓文, 馬拉加斯文, 馬拉地文, 馬拉雅拉姆文, 馬耳他文, 高棉文, 等語言的翻譯.

Copyright ©2025 I Love Translation. All reserved.

E-mail: